GRAFTMASTER 1012580-26

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-13 for GRAFTMASTER 1012580-26 manufactured by Abbott Vascular.

Event Text Entries

[183929546] The device was not returned for analysis. A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event. Additionally, a review of the complaint history identified no other incidents. The investigation was unable to determine a conclusive cause for the reported failure to advance; however, the subsequent treatment appears to be related to circumstances of the procedure. There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
Patient Sequence No: 1, Text Type: N, H10


[183929547] It was reported that a perforation was being treated; however, a 2. 8x26mm graftmaster rx covered stent failed to cross. The perforation was successfully treated with a 3. 0x15mm non-abbott stent. No additional information was provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2024168-2020-02429
MDR Report Key9832461
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-13
Date of Report2020-03-13
Date of Event2020-01-04
Date Mfgr Received2020-02-20
Device Manufacturer Date2018-12-05
Date Added to Maude2020-03-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. LINDSEY BELL
Manufacturer Street26531 YNEZ RD.
Manufacturer CityTEMECULA CA 925914628
Manufacturer CountryUS
Manufacturer Postal925914628
Manufacturer Phone9519143996
Manufacturer G1ABBOTT VASCULAR, REG # 3005718570 (P099)
Manufacturer StreetCASHEL ROAD
Manufacturer CityCLONMEL TIPPERARY
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGRAFTMASTER
Generic NameCORONARY STENT DELIVERY SYSTEM
Product CodeMAF
Date Received2020-03-13
Model Number1012580-26
Catalog Number1012580-26
Lot Number8120541
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerABBOTT VASCULAR
Manufacturer Address26531 YNEZ RD. TEMECULA CA 925914628 US 925914628


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.