MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2020-03-14 for 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV) 3100A 768901 manufactured by Vyaire Medical..
[184322869]
(b)(4). Vyaire field service representative (fsr) went onsite and found map and limit knobs were misadjusted. Fsr corrected the settings, performed circuit calibration and performance checkout. System passes but performance check is on low side. Performed 2k preventive maintenance and replaced 2 filters. Performed circuit calibration and performance check. Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803. 56 if additional information becomes available.
Patient Sequence No: 1, Text Type: N, H10
[184322870]
The customer reported that the 3100a map (mean airway pressure) is fluctuating without touching controls and the unit is failing performance check. There is no patient involvement in this reported event.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2021710-2020-11517 |
MDR Report Key | 9832730 |
Report Source | HEALTH PROFESSIONAL,USER FACI |
Date Received | 2020-03-14 |
Date of Report | 2020-02-13 |
Date of Event | 2020-02-13 |
Date Mfgr Received | 2020-02-13 |
Date Added to Maude | 2020-03-14 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MINDY FABER |
Manufacturer Street | 26125 N. RIVERWOODS BLVD. |
Manufacturer City | METTAWA, IL |
Manufacturer Country | US |
Manufacturer Phone | 7570116 |
Manufacturer G1 | VYAIRE MEDICAL INC. |
Manufacturer Street | 1100 BIRD CENTER DRIVE |
Manufacturer City | PALM SPRINGS, CA |
Manufacturer Country | US |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV) |
Generic Name | VENTILATOR, HIGH FREQUENCY |
Product Code | LSZ |
Date Received | 2020-03-14 |
Model Number | 3100A |
Catalog Number | 768901 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | VYAIRE MEDICAL. |
Manufacturer Address | 26125 N. RIVERWOODS BLVD. METTAWA, IL US |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-14 |