E300 LED CEILING MOUNTED DENTAL LIGHT SYSTEM P024617

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2020-03-13 for E300 LED CEILING MOUNTED DENTAL LIGHT SYSTEM P024617 manufactured by Engle Dental Systems, Llc.

Event Text Entries

[183894616] Ceiling light installed by 3rd party equipment dealer ((b)(6) dental services) fell from installation point while in use at (b)(6) dental, (b)(6). A dental assistant and pt caught the falling light mid-air preventing an injury. Upon investigation, key mounting instructions and parts were not used in the installation of the light, clearly instructions on setup and use were not followed. Factory directed service inspection was provided after notification of incident. During the inspection the findings were clear that the equipment was installed improperly. A warning was provided to both (b)(6) dental and (b)(6) dental services that the equipment was originally installed incorrectly and posed a hazard. A second factory directed installer went through the office inspecting equipment, retrieved the damaged light, and noted the improper installation of the damaged light along with another light still in use. Both the factory directed installer and engle dental have provided advanced dental a warning the equipment needs to be taken out of service or re-installed under proper supervision. Engle dental has instructed dental practice to remove all lights from practice use or have them properly installed by a trained and knowledgeable service tech. As of the report date, (b)(6) dental practice mgmt and drs have not complied with removal or proper installation warnings. Fda safety report id# (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5093757
MDR Report Key9834827
Date Received2020-03-13
Date of Report2020-03-11
Date of Event2020-02-18
Date Added to Maude2020-03-16
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameE300 LED CEILING MOUNTED DENTAL LIGHT SYSTEM
Generic NameLIGHT, OPERATING, DENTAL
Product CodeEAZ
Date Received2020-03-13
Returned To Mfg2020-02-28
Model NumberP024617
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerENGLE DENTAL SYSTEMS, LLC
Manufacturer AddressHILLSBORO OR 97124 US 97124


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.