EM 2400, LOAD CELL MODULE 2400L

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-16 for EM 2400, LOAD CELL MODULE 2400L manufactured by Baxter Healthcare Corportation.

Event Text Entries

[183502892] The device was received for evaluation. During visual examination, the cord is frayed with exposed wires with several internal conductors cut and exposed. The reported condition was verified. Due to the nature of the returned sample, no further testing could be performed. The cause of the condition could not be determined. To resolve the condition, the scale cable assembly will be replaced. A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[183502893] It was reported that the cord (load cell module) of an exacta mix automated compounding device were damaged. It was further reported that the load cell wires were frayed and exposed. This issue was noted during programming/setup. There was no patient involvement. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1416980-2020-01414
MDR Report Key9834871
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-16
Date of Report2020-03-16
Date of Event2020-02-18
Date Mfgr Received2020-02-18
Date Added to Maude2020-03-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPHARMACIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1ECM - ENGLEWOOD
Manufacturer Street14445 GRASSLANDS DR
Manufacturer CityENGLEWOOD CO 80112
Manufacturer CountryUS
Manufacturer Postal Code80112
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEM 2400, LOAD CELL MODULE
Generic NameSYSTEM/DEVICE, PHARMACY COMPOUNDING
Product CodeNEP
Date Received2020-03-16
Returned To Mfg2020-02-21
Model NumberNA
Catalog Number2400L
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE CORPORTATION
Manufacturer AddressDEERFIELD IL


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.