TACTICATH? QUARTZ CONTACT FORCE ABLATION CATHETER, 65MM PN-004 065 PN-004065

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-16 for TACTICATH? QUARTZ CONTACT FORCE ABLATION CATHETER, 65MM PN-004 065 PN-004065 manufactured by Venusa De Mexico S.a. De C.v..

Event Text Entries

[183516709] The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
Patient Sequence No: 1, Text Type: N, H10


[183516710] Related manufacturing ref: 9680001-2020-00015. During the procedure, there was no contact force reading on the catheter. The catheter was exchanged which did not resolve the issue and the procedure was cancelled.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9680001-2020-00014
MDR Report Key9834988
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-16
Date of Report2020-03-16
Date of Event2020-03-07
Date Mfgr Received2020-03-08
Date Added to Maude2020-03-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEPHANIE O' SULLIVAN
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1VENUSA DE MEXICO S.A. DE C.V.
Manufacturer StreetCALLE HERTZ 1525-6 PARQUE INDUSTRIAL J. BERMUDEZ
Manufacturer CityCIUDAD JUAREZ, CHIHUAHUA 32470
Manufacturer CountryMX
Manufacturer Postal Code32470
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTACTICATH? QUARTZ CONTACT FORCE ABLATION CATHETER, 65MM
Generic NameCATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Product CodeOAE
Date Received2020-03-16
Model NumberPN-004 065
Catalog NumberPN-004065
Lot Number6884659
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerVENUSA DE MEXICO S.A. DE C.V.
Manufacturer AddressCALLE HERTZ 1525-6 PARQUE INDUSTRIAL J. BERMUDEZ CIUDAD JUAREZ, CHIHUAHUA 32470 MX 32470


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-03-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.