[183568402] The devices were described with karl storz brand name but without model number and there is no contact information for us to follow up with the customer. We suspect they were our light cable item 495udl (bifurcated fiber optic, cable) connected to the ureteral probe 496u (ureter probe illuminated) and to the integra luxtec portable light source. We suspect the light cable and probe connector was in contact with the patient's thigh and resulted in the burn, we therefore file this event on the light cable. Our ifu 97000213 v3. 0 (11-2017) warns the user not to rest the light cable on the patient for risk of injury.
Patient Sequence No: 1, Text Type: N, H10


[183568403] We received medwatch report mw5092762, it stated a patient sustained a burn on the left thigh after ureteral lighted stents connected to a portable light box were taped to the thigh.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610617-2020-00035
MDR Report Key9835664
Report SourceUSER FACILITY
Date Received2020-03-16
Date of Report2020-03-11
Date of Event2020-01-29
Date Mfgr Received2020-02-25
Device Manufacturer Date2016-12-01
Date Added to Maude2020-03-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMRS. SUSIE CHEN
Manufacturer Street2151 E. GRAND AVENUE
Manufacturer CityEL SEGUNDO, CA
Manufacturer CountryUS
Manufacturer Phone2188201
Manufacturer G1KARL STORZ SE & CO.KG
Manufacturer StreetDR.-KARL-STORZ -STRASSE 34 78532
Manufacturer CityTUTTLINGEN, GM,
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameFIBER OPTIC LIGHT CABLE
Generic NameLIGHT CABLE BIFURCATED AT TELESCOPE END 3.5 X 300CM
Product CodeFCW
Date Received2020-03-16
Model Number495UDL
Catalog Number495UDL
Lot NumberZT
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerKARL STORZ SE & CO. KG
Manufacturer AddressDR.-KARL-STORZ-STRASSE 34 78532 TUTTLINGEN, GM, US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-16

© 2026 FDA.report