4MM FEMORAL AIMER 234-020-044 0234020044

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-03-16 for 4MM FEMORAL AIMER 234-020-044 0234020044 manufactured by Stryker Endoscopy-san Jose.

Event Text Entries

[186349512] Alleged failure: the 2 drill bit were returned because after the first use (ok-no problem) it rubs when passing (+overheating) through the guide and then it blocks. The failure alleged in the complaint record was not confirmed/duplicated during the product investigation. The probable root causes could be: excessive force applied on device or rough handling during reprocessing. The product was returned for investigation and the failure mode will be monitored for future reoccurrence. The device manufacturer date is not known.
Patient Sequence No: 1, Text Type: N, H10


[186349513] Per investigation results, the tip of the device was broken off. Per the event description, there is no alleged patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0002936485-2020-00125
MDR Report Key9835924
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-03-16
Date of Report2020-03-16
Date of Event2019-11-20
Date Mfgr Received2020-02-18
Date Added to Maude2020-03-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. ANDREA ZENERE
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal95138
Manufacturer Phone4087542000
Manufacturer G1STRYKER ENDOSCOPY-SAN JOSE
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal Code95138
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name4MM FEMORAL AIMER
Generic NameACCESSORIES, ARTHROSCOPIC
Product CodeNBH
Date Received2020-03-16
Returned To Mfg2020-02-04
Model Number234-020-044
Catalog Number0234020044
Lot Number623710
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER ENDOSCOPY-SAN JOSE
Manufacturer Address5900 OPTICAL COURT SAN JOSE CA 95138 US 95138


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-16