PLATE CUTTER FOR 1.3MM/1.5MM AND 2.0MM PLATES 03.130.271

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-03-16 for PLATE CUTTER FOR 1.3MM/1.5MM AND 2.0MM PLATES 03.130.271 manufactured by Wrights Lane Synthes Usa Products Llc.

Event Text Entries

[188831798] Reporter is a synthes employee. Part: 03. 130. 271, lot: t155241, manufacturing site: tuttlingen, release to warehouse date: january 23, 2019. A review of the device history records was performed for the finished device lot number and no relevant nonconformance's were found. The final quality release criteria were met before this batch was released for distribution. Visual inspection: the plate cutter for 1. 3mm/1. 5mm and 2. 0mm plates was received at us customer quality (cq). Upon visual inspection, one of the screws is loose since the stakes are broken. The silicone plate holder inserts are also missing. Dimensional inspection: no dimensional inspection was performed due to post-manufacturing damage. Document/specification review: current and manufactured revisions were reviewed. No design issues or discrepancies were identified. Investigation conclusion: this complaint is confirmed as a screw is loose since the stakes are broken and the plate holder inserts are missing. No definitive root cause could be determined based on the provided information. No new, unique, or different patient harms were identified as a result of this evaluation. There was no indication that a design or manufacturing issue contributed to the complaint. No design issues were observed during the document/specification review. Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed. Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities. Device was used for treatment, not diagnosis. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
Patient Sequence No: 1, Text Type: N, H10


[188831799] It was reported that during routine incoming inspection of a loaner set on an unknown date, it was observed that the plate cutter was damaged. There was no known patient or hospital involvement. Upon further investigation, it was determined that the device was broken. This report is for a plate cutter. This is report 1 of 1 for (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2939274-2020-01294
MDR Report Key9836651
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-03-16
Date of Report2020-01-31
Date Mfgr Received2020-02-21
Device Manufacturer Date2019-01-23
Date Added to Maude2020-03-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKARA DITTY-BOVARD
Manufacturer Street1302 WRIGHTS LANE EAST
Manufacturer CityWEST CHESTER PA 19380
Manufacturer CountryUS
Manufacturer Postal19380
Manufacturer Phone6103142063
Manufacturer G1SYNTHES TUTTLINGEN
Manufacturer StreetUNTER HASSLEN 5
Manufacturer CityTUTTLINGEN 78532
Manufacturer CountryGM
Manufacturer Postal Code78532
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePLATE CUTTER FOR 1.3MM/1.5MM AND 2.0MM PLATES
Generic NameINSTRUMENT, CUTTING, ORTHOPAEDIC
Product CodeHTZ
Date Received2020-03-16
Returned To Mfg2020-02-11
Model Number03.130.271
Catalog Number03.130.271
Lot NumberT155241
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerWRIGHTS LANE SYNTHES USA PRODUCTS LLC
Manufacturer Address1302 WRIGHTS LANE EAST WEST CHESTER PA 19380 US 19380


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-16

© 2026 FDA.report