ARCTIC FRONT ADVANCE PRO? CARDIAC CRYOABLATION CATHETER AFAPRO28

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-16 for ARCTIC FRONT ADVANCE PRO? CARDIAC CRYOABLATION CATHETER AFAPRO28 manufactured by Medtronic Cryocath Lp.

Event Text Entries

[183567111] Product event summary: the data files were returned and analyzed. The files showed at least 12 applications were performed with a non-returned balloon catheter afapro28 with lot number 98927 on the date of the event. The pressure and temperature did not follow the preset value. A clinical issue (phrenic nerve palsy (pnp)) occurred during procedure. In conclusion, the balloon catheter was not returned for investigation. There is no indication of relation of adverse event to the performance and malfunction of the product. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[183567112] It was reported that during a cryo ablation procedure, the activity of the diaphragm weakened indicating phrenic nerve palsy (pnp). The double stop technique was performed. The case was completed with cryo. The patient was asymptomatic, but did not fully recover. No further patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002648230-2020-00161
MDR Report Key9836700
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-16
Date of Report2020-03-16
Date of Event2020-02-25
Date Mfgr Received2020-02-25
Device Manufacturer Date2019-09-24
Date Added to Maude2020-03-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1MEDTRONIC CRYOCATH LP
Manufacturer Street9000 AUTOROUTE TRANSCANADIENNE
Manufacturer CityPOINTE-CLAIRE,QC H9R 5Z8
Manufacturer CountryCA
Manufacturer Postal CodeH9R 5Z8
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARCTIC FRONT ADVANCE PRO? CARDIAC CRYOABLATION CATHETER
Generic NamePERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION
Product CodeOAE
Date Received2020-03-16
Model NumberAFAPRO28
Catalog NumberAFAPRO28
Lot Number98927
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC CRYOCATH LP
Manufacturer Address9000 AUTOROUTE TRANSCANADIENNE POINTE-CLAIRE,QC H9R 5Z8 CA H9R 5Z8


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-03-16