FOGARTY THRU-LUMEN EMBOLECTOMY CATHETER 12TLW405F35

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2020-03-16 for FOGARTY THRU-LUMEN EMBOLECTOMY CATHETER 12TLW405F35 manufactured by Edwards Lifesciences Pr.

Event Text Entries

[186023436] The device was not returned for evaluation due to hospital policy. However, the customer will send a photograph for evaluation purposes. A supplemental report will be forthcoming with any additional information. A review of the manufacturing records indicated that the product met specifications upon release. Udi# ((b)(4) # mw5092368.
Patient Sequence No: 1, Text Type: N, H10


[186023437] It was reported that during embolization procedure on a (b)(6) female patient for an av fistula, the balloon on the catheter ruptured while using a 5 fr embolectomy catheter. When the catheter was removed from the vascular site, the balloon was no longer attached. No apparent complications reported. Further follow up with the customer stated the device will not be sent but they are willing to provide photos. Edwards lifesciences was notified of the event by a (b)(4) report.
Patient Sequence No: 1, Text Type: D, B5


[186023729] A photo provided by the customer was reviewed and the reported event of "balloon on the catheter ruptured" was confirmed from the images. Two images were reviewed. Images showed a fogarty through lumen catheter in a biohazard bag. Red spots appeared to be blood were observed on the catheter tip, the through lumen hub and the gate valve. The balloon appeared to be ruptured between the balloon bushings. A white spot, that appeared to be catheter body was observed from under the ruptured balloon, between the balloon bushing. The images were not clear to confirm if the balloon edges matched at the ruptured location. The reported event of detached balloon was not able to be confirmed from the images as the images were not clear. An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions. The fogarty thru-lumen embolectomy catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the arterial system. The thru-lumen embolectomy catheter is not recommended for the removal of fibrous, adherent or calcified material such as chronic clot or atherosclerotic plaque. The catheter is not designed to withstand the additional pull force needed to remove these materials. The ifu for the product contains the following warning: balloon rupture and catheter separation as a result of excessive pull force applied to remove adherent material are the most frequent causes of reported failures. The possibility of balloon rupture must be taken into account when considering the risks involved in any embolectomy procedure. To minimize the risk of vessel damage, balloon rupture, or tip detachment, do not exceed the maximum recommended inflation volume and pull force for each size catheter. Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2015691-2020-10982
MDR Report Key9838144
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2020-03-16
Date of Report2020-02-24
Date of Event2020-01-07
Date Mfgr Received2020-03-16
Device Manufacturer Date2019-10-25
Date Added to Maude2020-03-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS LYNN THOMAS
Manufacturer Street1 EDWARDS WAY
Manufacturer CityIRVINE CA 92614
Manufacturer CountryUS
Manufacturer Postal92614
Manufacturer Phone9497564386
Manufacturer G1EDWARDS LIFESCIENCES PR
Manufacturer StreetSTATE RD INDUS PK 402 KM 1.4
Manufacturer CityANASCO PR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameFOGARTY THRU-LUMEN EMBOLECTOMY CATHETER
Generic NameCATHETER, EMBOLECTOMY
Product CodeDXE
Date Received2020-03-16
Model Number12TLW405F35
Catalog Number12TLW405F35
Lot Number62667400
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerEDWARDS LIFESCIENCES PR
Manufacturer AddressSTATE RD INDUS PK 402 KM 1.4 ANASCO PR