SURG-O-VAC ASPIRATION TIP DENTAL 077125 SURG-O-VAC 4MM OPENING 077125

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2020-03-16 for SURG-O-VAC ASPIRATION TIP DENTAL 077125 SURG-O-VAC 4MM OPENING 077125 manufactured by Young Dental Manufacturing Co I, Llc..

Event Text Entries

[183915338] Our facility's dental clinic uses young dental surg-o-vac aspiration tips, which we have been sterilizing as the package label says autoclavable and letter received 12/03/2012 says autoclavable up to 135c. On (b)(6) 2020 i emailed the company for more explicit sterilization parameters (i. E, gravity versus prevacuum cycle and exposure time). Company replied: "the memo forwarded with the customer injury is several years old and out of date. We do not have a validated sterilization protocol and so have since removed any info from the product's labeling regarding sterilization or the use of an autoclave in a sterilization process. There have been no other changes to the material or mfg process after other than the change to the labeling /ifu. " i email-replied that dental supply companies are still advertising surg-o-vac tips as autoclavable. (b)(6). After receiving no response to date ((b)(6) 2020) i called the company and spoke to the same customer service agent who emailed me. She said if she hadn't emailed me it was because young's quality and regulatory dept hadn't gotten back to her. I asked if i could speak to someone in the quality and regulatory dept agent said no. I asked if they had phones and she said no. Product is still being advertised as autoclavable and product packaging still says autoclavable, but company has no validation for sterilization. Fda safety report id# (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5093776
MDR Report Key9840196
Date Received2020-03-16
Date of Report2020-03-13
Date of Event2020-03-13
Date Added to Maude2020-03-17
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSURG-O-VAC ASPIRATION TIP DENTAL
Generic NameEVACUATOR, ORAL CAVITY
Product CodeEHZ
Date Received2020-03-16
Model Number077125 SURG-O-VAC 4MM OPENING
Catalog Number077125
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by MfgrI
Device Sequence No1
Device Event Key0
ManufacturerYOUNG DENTAL MANUFACTURING CO I, LLC.


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-03-16