INGENIA 1.5T 781396

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-17 for INGENIA 1.5T 781396 manufactured by Philips Electronics Nederland B.v..

Event Text Entries

[183896000] The patient heating form was send today to customer. Actually the event date is unknown, waiting information from customer. Translation of the attached email: good morning, (b)(6), thank you for the confirmation of the software version without any connection, (b)(6) alerted me to the occurrence of two very close cases of burns on the left side of the patient, in the normal direction. 1st case of proven burns on the breast, obese patient, push button (shirt) in contact with the mri hood and the breast. Mri in progress it has taken steps to isolate the patient's skin with the mri hood. No antenna elements (flexible antenna) are in contact with the patient. I am asking for your opinion. Can we rule out the hypothesis of heating at the level of a gradient coil? Of another component? Do you have any internal probes? A monitoring of the temperature or power consumption of a component located at this location? How could we prove the absence/presence of internal heating on mri? Your elements should allow us to make the right hypotheses. Thank you for your input, (b)(6).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003768277-2020-01003
MDR Report Key9841492
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-17
Date of Event2020-01-30
Date Facility Aware2020-02-27
Date Mfgr Received2020-02-27
Date Added to Maude2020-03-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. DUSTY LEPPERT
Manufacturer StreetVEENPLUIS 6
Manufacturer CityBEST 5684PC
Manufacturer CountryNL
Manufacturer Postal5684 PC
Manufacturer G1PHILIPS ELECTRONICS NEDERLAND B.V.
Manufacturer StreetVEENPLUIS 6
Manufacturer CityBEST 5684PC
Manufacturer CountryNL
Manufacturer Postal Code5684 PC
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINGENIA 1.5T
Generic NameSYSTEM, NUCLEAR MAGNETIC RESONANCE IMAGING
Product CodeLNH
Date Received2020-03-17
Model NumberINGENIA 1.5T
Catalog Number781396
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPHILIPS ELECTRONICS NEDERLAND B.V.
Manufacturer AddressVEENPLUIS 6 BEST 5684PC NL 5684 PC


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-03-17

© 2024 FDA.report