JETSTREAM XC 45007

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-17 for JETSTREAM XC 45007 manufactured by Boston Scientific Corporation.

Event Text Entries

[183878360] (b)(6). Device evaluated by manufacturer: returned product consisted of a jetstream xc-2. 1 atherectomy catheter. Visual examination noticed that the infusion, electrical and baton were cut from the pod. With this damage functionality of the device could not be confirmed. It was noticed that there was shaft damage in the form of buckling/kinks located 4cm and 80. 5cm from the tip. Microscope examination noticed that the infusion line had burst proximal the buckling/kinks. The location of the burst infusion line was approximately 89cm to 91cm from the tip. Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
Patient Sequence No: 1, Text Type: N, H10


[183878361] It was reported that the device fell apart and leaked. A 2. 1mm jetstream xc catheter was selected for an atherectomy procedure in the severely calcified right superficial femoral artery (sfa). The physician noticed that the shaft of the catheter started to fall apart and leak towards the end of the procedure. The procedure was completed successfully with this device. There were no patient complications reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134265-2020-03235
MDR Report Key9841771
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-17
Date of Report2020-03-17
Date of Event2020-02-25
Date Mfgr Received2020-02-25
Device Manufacturer Date2019-07-08
Date Added to Maude2020-03-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAY JOHNSON
Manufacturer StreetTWO SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal55311
Manufacturer Phone7634942574
Manufacturer G1BOSTON SCIENTIFIC CORK LIMITED
Manufacturer StreetMODEL FARM ROAD
Manufacturer CityCORK
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameJETSTREAM XC
Generic NameCATHETER, PERIPHERAL, ATHERECTOMY
Product CodeMCW
Date Received2020-03-17
Returned To Mfg2020-03-10
Model Number45007
Catalog Number45007
Lot Number0024066237
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer AddressTWO SCIMED PLACE MAPLE GROVE MN 55311 US 55311


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-17

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.