MINIMAL ACCESS ATTACHMENT MA-D20

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-03-17 for MINIMAL ACCESS ATTACHMENT MA-D20 manufactured by Depuy Synthes Products Llc.

Event Text Entries

[186736467] Device was used for treatment, not diagnosis. If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate. Device evaluation: the actual device was returned for evaluation. The device was evaluated, and the reported condition was not confirmed. Therefore, an assignable root cause was not determined. However, during repairs, a visual and functional assessment was performed which determined that the bearings were damaged and the mid and back assemblies were contaminated. It was determined that the issues were consistent with improper maintenance. The assignable root cause was determined to be traced to maintenance, which is improper maintenance interval. If additional information should become available, a supplemental medwatch report will be submitted accordingly.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1045834-2020-00479
MDR Report Key9841826
Report SourceUSER FACILITY
Date Received2020-03-17
Date of Report2020-03-06
Date Mfgr Received2020-03-06
Device Manufacturer Date2014-09-04
Date Added to Maude2020-03-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKARA DITTY-BOVARD
Manufacturer Street4500 RIVERSIDE DRIVE
Manufacturer CityPALM BEACH GARDENS FL 33410
Manufacturer CountryUS
Manufacturer Postal33410
Manufacturer Phone6103142063
Manufacturer G1DEPUY SYNTHES PRODUCTS LLC
Manufacturer Street4500 RIVERSIDE DRIVE
Manufacturer CityPALM BEACH GARDENS FL 33410
Manufacturer CountryUS
Manufacturer Postal Code33410
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMINIMAL ACCESS ATTACHMENT
Generic NameDRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Product CodeHBE
Date Received2020-03-17
Model NumberMA-D20
Catalog NumberMA-D20
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerDEPUY SYNTHES PRODUCTS LLC
Manufacturer Address4500 RIVERSIDE DRIVE PALM BEACH GARDENS FL 33410 US 33410


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-17

© 2026 FDA.report