RF DISPOSABLE GROUNDING PAD

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-17 for RF DISPOSABLE GROUNDING PAD manufactured by Neurotherm, Inc.

Event Text Entries

[187167705] Investigation results will be provided in a subsequent submission. Further information was requested but not received.
Patient Sequence No: 1, Text Type: N, H10


[187167706] One week following a radio frequency ablation (rfa), the patient returned to the surgery center with a burn on his leg from the grounding pad. During the procedure, the skin was not prepped and the grounding pad was placed on a hairy part of the leg for a cervical rfa. Attempts to ablate were unsuccessful. The grounding pad was noted to be too far away and was moved to the patients upper back to complete the procedure with no issues. The patient returned to the surgery center one week post procedure with 2nd degree burns on his leg. The generator has been used since without any issues.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002953813-2020-00011
MDR Report Key9841902
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-17
Date of Report2020-03-26
Date of Event2020-02-26
Date Mfgr Received2020-03-04
Date Added to Maude2020-03-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEPHANIE O' SULLIVAN
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1NEUROTHERM, INC
Manufacturer Street600 RESEARCH DRIVE STE 1
Manufacturer CityWILMINGTON MA 01887
Manufacturer CountryUS
Manufacturer Postal Code01887
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRF DISPOSABLE GROUNDING PAD
Generic NameELECTROSURGICAL RETURN ELECTRODE
Product CodeGXD
Date Received2020-03-17
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerNEUROTHERM, INC
Manufacturer Address600 RESEARCH DRIVE STE 1 WILMINGTON MA 01887 US 01887


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-03-17