IDEAL SUTGRASPER 60 DEG *EA 251723

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign report with the FDA on 2020-03-17 for IDEAL SUTGRASPER 60 DEG *EA 251723 manufactured by Depuy Mitek Llc Us.

Event Text Entries

[188645748] Product complaint # (b)(4). If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate. Udi: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[188645749] It was reported by affiliate via personal interaction that during procedure the surgeon retracted the ideal sutgrasper 60 deg ea. In closed position out of the joint, then the wire broke and was lost. The procedure was completed with 30 minute delay. No patient consequences reported. The device is available to be returned for evaluation. Additional information provided by the affiliate reported the broken wire was a part of the grasper and not the implant. It was also reported the broken wire was possibly left in the patient and additional information regarding the status of the patient could not be provided. The affiliate reported the device will not be returning for evaluation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1221934-2020-00875
MDR Report Key9842609
Report SourceFOREIGN
Date Received2020-03-17
Date of Report2020-03-03
Date of Event2020-03-03
Date Mfgr Received2020-03-25
Device Manufacturer Date2019-08-09
Date Added to Maude2020-03-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKARA DITTY-BOVARD
Manufacturer Street325 PARAMOUNT DRIVE
Manufacturer CityRAYNHAM MA 02767
Manufacturer CountryUS
Manufacturer Postal02767
Manufacturer Phone6103142063
Manufacturer G1MEDOS INT MITEK
Manufacturer StreetCHEMIN BLANC 38
Manufacturer CityLE LOCLE 02400
Manufacturer CountrySZ
Manufacturer Postal Code02400
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIDEAL SUTGRASPER 60 DEG *EA
Generic NameSUTURE/NEEDLE PASSER, REUSABLE
Product CodeNBH
Date Received2020-03-17
Model Number251723
Catalog Number251723
Lot Number5L45473
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerDEPUY MITEK LLC US
Manufacturer Address325 PARAMOUNT DRIVE RAYNHAM MA 02767 US 02767


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-17