AMS SPECTRA CONCEALABLE PENILE PROSTHESIS 720074-03

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-03-17 for AMS SPECTRA CONCEALABLE PENILE PROSTHESIS 720074-03 manufactured by Boston Scientific Corporation.

Event Text Entries

[183751629] It was reported that the patient had a removal of his inflatable penile prosthesis (ipp) due to scrotal pain. The ipp was replaced with a spectra malleable penile prosthesis (spp). During the case, when the doctor was about to implant the spp on the right side of the corpora, he noticed that there was missing part/piece of coating on the outside of the device (visible pieces of metal inside). Another device with different lot number was used to complete the procedure (one lot number on one side of the penis and another lot number on the other side). One of the device/rod (720074-03/120414004 left cylinder) was defective, while (720074-03/ 1000132285 right cylinder) was not defective. The procedure was completed and no patient complications were reported in relation to this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2183959-2020-01320
MDR Report Key9842617
Report SourceHEALTH PROFESSIONAL
Date Received2020-03-17
Date of Report2020-03-27
Date of Event2020-02-25
Date Mfgr Received2020-03-20
Device Manufacturer Date2016-02-18
Date Added to Maude2020-03-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactALYSON HARRIS
Manufacturer Street10700 BREN ROAD W
Manufacturer CityMINNETONKA MN 55343
Manufacturer CountryUS
Manufacturer Postal55343
Manufacturer Phone4089353452
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer Street10700 BREN ROAD W
Manufacturer CityMINNETONKA MN 55343
Manufacturer CountryUS
Manufacturer Postal Code55343
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAMS SPECTRA CONCEALABLE PENILE PROSTHESIS
Generic NamePROSTHESIS PENILE
Product CodeFAE
Date Received2020-03-17
Returned To Mfg2020-03-05
Model Number720074-03
Catalog Number720074-03
Lot Number0120414004
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer Address10700 BREN ROAD W MINNETONKA MN 55343 US 55343


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-17