COOK UROCOGICAL STENT N/A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1993-08-02 for COOK UROCOGICAL STENT N/A manufactured by Cook Medical.

Event Text Entries

[7523] To or for removal of bilateral renal stents. Left ureteral stent broke off during removal attempt with alligator forceps. Approximately 2/3 of stent remains in patient. Right ureteral stent removed without incident. Stent implanted 5/27/93 a urological stent, multi-length silicone. Set #1491-358 #sfr. 22-23cm. Removal attempted 6/21/93. Device labeled for single use. Patient medical status prior to event: satisfactory condition. There was not multiple patient involvement. Invalid data - on device service/maintenance. No data - regarding date last serviced. Service provided by: invalid data. Invalid data - service records availability. No imminent hazard to public health claimed. Device used as labeled/intended. Device was evaluated after the event. Method of evaluation: actual device involved in incident was evaluated, visual examination. Results of evaluation: none or unknown, invalid data. Conclusion: device failure occurred and was related to event, device failure directly caused event. Certainty of device as cause of or contributor to event: yes. Corrective actions: device permanently removed from service. The device was not destroyed/disposed of.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9843
MDR Report Key9843
Date Received1993-08-02
Date of Report1993-06-25
Date of Event1993-06-21
Date Facility Aware1993-06-21
Report Date1993-06-25
Date Reported to Mfgr1993-06-21
Date Added to Maude1995-02-01
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameCOOK UROCOGICAL STENT
Generic NameSTENT
Product CodeMER
Date Received1993-08-02
Model NumberN/A
Catalog NumberN/A
Lot Number1491/358
ID NumberN/A
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityY
Implant FlagY
Device Sequence No1
Device Event Key9495
ManufacturerCOOK MEDICAL


Patients

Patient NumberTreatmentOutcomeDate
101. Other 1993-08-02