RAPIDPOINT 500 BLOOD GAS ANALYZER 10492730

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2020-03-17 for RAPIDPOINT 500 BLOOD GAS ANALYZER 10492730 manufactured by Siemens Healthcare Diagnostics.

Event Text Entries

[183973660] This issue has been escalated to software engineering for evaluation. The trace log provided by the customer had the incorrect time frame. Siemens has requested the trace log with the correct time frame. The cause of this event is unknown.
Patient Sequence No: 1, Text Type: N, H10


[183973661] The customer reported they ran a patient on the rp 500 and the user confirms the patient id was scanned correctly, all results on the patient's chart are correct, the printouts from the analyzer are correct and the rapidcomm results are correct. When the user recalled the patient list, several results from the previous 24 hours were all overwritten with one, different patient id. There was no report of injury due to the event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002637618-2020-00012
MDR Report Key9844902
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2020-03-17
Date of Report2020-03-17
Date of Event2020-02-17
Date Mfgr Received2020-02-24
Date Added to Maude2020-03-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactFELIX AKINRINOLA
Manufacturer Street2 EDGEWATER DRIVE
Manufacturer CityNORWOOD, MA
Manufacturer CountryUS
Manufacturer Phone7052212
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
Manufacturer StreetNORTHERN ROAD CHILTON INDUSTRIAL ESTATE
Manufacturer CitySUDBURY, CO102XQ
Manufacturer CountryUK
Manufacturer Postal CodeCO10 2XQ
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRAPIDPOINT 500 BLOOD GAS ANALYZER
Generic NameRP 500
Product CodeCHL
Date Received2020-03-17
Model Number10492730
Catalog Number10492730
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer Address511 BENEDICT AVENUE TARRYTOWN, NY US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-17

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.