BD VACUTAINER? Z (NO ADDITIVE) PLUS URINE TUBE 364915

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,other,user facility report with the FDA on 2020-03-17 for BD VACUTAINER? Z (NO ADDITIVE) PLUS URINE TUBE 364915 manufactured by Becton, Dickinson And Company (bd).

Event Text Entries

[186303508] Medical device expiration date: unknown. Device manufacture date: unknown. A device evaluation and/or device history review is anticipated, but is not complete. Upon completion, a supplemental report will be filed.
Patient Sequence No: 1, Text Type: N, H10


[186303509] It was reported that bd vacutainer? Z (no additive) plus urine tube contained foreign matter. The following information was provided by the initial reporter: "plastic flakes when tubes go through tempus. Customer has a tempus pneumatic tube system for sending tubes 1 by 1 on air pressure since 2017. Since 1 year they noticed particulates being shed at the fan of the system. The fan is above the ceiling so the particulates would collect themselves on the ceiling. (see pictures for situation and particulates). Suspicion is that the particulates is plastic from the tubes. The customer has asked the supplier of the transport system first but also got us involved now. Hypothesis, since the homogeneity of the particulates, that it might be the big 16x100 urine tube they use, so this complaint is filed on that product number but not sure it is this product and not able to pinpoint a lot number since a lot tubes are involved / sent over the pneumatic system over the years. We're looking with the customers for more investigations. "
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9617032-2020-00255
MDR Report Key9845081
Report SourceFOREIGN,OTHER,USER FACILITY
Date Received2020-03-17
Date of Report2020-03-18
Date of Event2020-02-25
Date Mfgr Received2020-02-25
Date Added to Maude2020-03-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBDX BRETT WILKO
Manufacturer Street9450 SOUTH STATE STREET
Manufacturer CitySANDY UT 84070
Manufacturer CountryUS
Manufacturer Postal84070
Manufacturer Phone8015652341
Manufacturer G1BECTON, DICKINSON AND COMPANY (BD)
Manufacturer StreetBELLIVER WAY BELLIVER INDUSTRIAL ESTATE
Manufacturer CityPLYMOUTH
Manufacturer CountryUK
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBD VACUTAINER? Z (NO ADDITIVE) PLUS URINE TUBE
Generic NameTRANSPORT CULTURE MEDIUM
Product CodeJSL
Date Received2020-03-17
Catalog Number364915
Lot NumberUNKNOWN
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBECTON, DICKINSON AND COMPANY (BD)
Manufacturer AddressBELLIVER WAY BELLIVER INDUSTRIAL ESTATE PLYMOUTH UK


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-03-17

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.