FLOSEAL 1503278

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-17 for FLOSEAL 1503278 manufactured by Baxter Healthcare Corporation.

Event Text Entries

[183858210] Two actual samples were received for evaluation. A visual inspection was performed under the microscope which revealed a reddish colored particulate matter in the needle syringe of the samples. The calcium chloride vial and the reddish pm inside the needle syringe were inspected under a microscope. The pm inside the needle syringe was of similar color, texture and shape as the missing segment from the calcium chloride vial stopper puncture site. The reported problem was verified. The cause of the particulate matter was found to be due to stopper coring; however the cause of coring was unable to be determined. A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[183858211] During evaluation a reddish colored particulate matter was observed in the needle syringe of two floseals. There was no patient involvement. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1416980-2020-01491
MDR Report Key9845140
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-17
Date of Report2020-03-17
Date of Event2020-01-10
Date Mfgr Received2020-02-27
Date Added to Maude2020-03-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1BAXTER HEALTHCARE - HAYWARD
Manufacturer Street2024 W WINTON AVE
Manufacturer CityHAYWARD CA 94545
Manufacturer CountryUS
Manufacturer Postal Code94545
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameFLOSEAL
Generic NameAGENT, ABSORBABLE HEMOSTATIC, COLLAGEN BASED
Product CodeLMF
Date Received2020-03-17
Model NumberNA
Catalog Number1503278
Lot NumberHA190885
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE CORPORATION
Manufacturer AddressDEERFIELD IL


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-17

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.