ZOLL IVTM THERMOGARD XP 8700-0650

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2020-03-17 for ZOLL IVTM THERMOGARD XP 8700-0650 manufactured by Zoll Circulation.

Event Text Entries

[188240706] The customer's reported complaint of the thermogard console displayed tcmid:06 (ce post failure) was confirmed during the archive review but not during functional testing. Additionally, the console failed with tcmid:02 (cooling engine danfoss error) error during the initial functional testing, as a characteristic of the failed cooling engine as a result of wear and tear. The defective cooling engine was replaced to remedy the customer reported complaint. Upon visual inspection, the cooling engine compressor motor was noted leaking, thus confirming a failed cooling engine. As part of routine service during testing, the console was examined and the well gaskets were noted discolored and white rollers were worn out, the casters and screws on top lid hinge were loose and cover deck xp and front cover were cracked, unrelated to the reported complaint. The thermogard console is a reusable device and was manufactured in september 2011 and is more than 9 years old. The console has exceeded its expected service life of 5 years. Therefore, this type of physical damages found during the visual inspection are characteristics of normal wear and tear for the life of the device. A review of the archive data shows multiple tcmid:06 error messages, thus confirming the customer reported complaint. Following the repair, the thermogard console passed all the testing with no issue or faults observed. All tests and readings are within the specified limits and functioned as intended. Historical complaints were reviewed for service information related to the reported complaint and there was no similar complaint reported for the thermogard console with serial number (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[188240707] During a routine check, the thermogard console displayed tcmid:06 (ce post failure) error message. No patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3010617000-2020-00272
MDR Report Key9845274
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2020-03-17
Date of Report2020-03-17
Date of Event2020-02-21
Date Mfgr Received2020-02-21
Device Manufacturer Date2011-09-02
Date Added to Maude2020-03-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS KIMTHOA NGUYEN
Manufacturer Street2000 RINGWOOD AVE,
Manufacturer CitySAN JOSE, CA
Manufacturer CountryUS
Manufacturer G1ZOLL CIRCULATION
Manufacturer Street2000 RINGWOOD AVE.
Manufacturer CitySAN JOSE, CA
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameZOLL IVTM THERMOGARD XP
Generic NameTHERMAL REGULATING SYSTEM
Product CodeNCX
Date Received2020-03-17
Returned To Mfg2020-03-10
Model Number8700-0650
Catalog Number8700-0650
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerZOLL CIRCULATION
Manufacturer Address2000 RINGWOOD AVE. SAN JOSE, CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-17

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.