PERCEPTA QUAD CRT-P MRI SURESCAN W4TR01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,consum report with the FDA on 2020-03-17 for PERCEPTA QUAD CRT-P MRI SURESCAN W4TR01 manufactured by Medtronic Puerto Rico Operations Co..

Event Text Entries

[183826649] Concomitant medical products: 459888, lead, implanted: (b)(6)2019. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[183826650] It was reported that the cardiac resynchronization therapy pacemaker (crt-p) exhibited an unexpected decrease in longevity, possibly due to elevated right ventricular (rv) lead pacing thresholds and high left ventricular (lv) lead pacing thresholds that were noted to have been increasing over time. It was unknown if the elevated rv thresholds were due to patient physiology versus a product performance issue but the lv thresholds were believed to be due to anatomical placement or physiological conditions. Both leads were reprogrammed to reduce output from the crt-p. The crt-p and leads remain in use. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


[186356222] Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed. Analysis of the device memory showed the battery is approaching the indicator signifying the time for device replacement. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[188943532] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004209178-2020-05572
MDR Report Key9845775
Report SourceCOMPANY REPRESENTATIVE,CONSUM
Date Received2020-03-17
Date of Report2020-03-23
Date of Event2020-01-08
Date Mfgr Received2020-03-17
Device Manufacturer Date2018-12-05
Date Added to Maude2020-03-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1MEDTRONIC PUERTO RICO OPERATIONS CO.
Manufacturer StreetROAD 31, KM. 24, HM 4 CEIBA NORTE INDUSTRIAL PARK
Manufacturer CityJUNCOS PR 00777
Manufacturer CountryUS
Manufacturer Postal Code00777
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePERCEPTA QUAD CRT-P MRI SURESCAN
Generic NamePULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION
Product CodeNKE
Date Received2020-03-17
Model NumberW4TR01
Catalog NumberW4TR01
OperatorLAY USER/PATIENT
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC PUERTO RICO OPERATIONS CO.
Manufacturer AddressROAD 31, KM. 24, HM 4 CEIBA NORTE INDUSTRIAL PARK JUNCOS PR 00777 US 00777


Patients

Patient NumberTreatmentOutcomeDate
150761. Required No Informationntervention 2020-03-17

© 2026 FDA.report