[183937791] Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation. Therefore, we are unable to determine the definitive cause of the reported event. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[183937792] It was reported that the patient underwent tightening of set screws and broken off per usual specifications at ilium due to neuromuscular scoliosis. Post-op, the set screw had come loose from the ballast screws. There were no patient complications as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


[188149499] Radiographic image review result: post of x-ray for lumbar/pelvic fixation provided. Unclear levels instrumented above this. Fusion status is unknown. Both of the iliac set screws were out of the screw heads. Unable to evaluate rod contour or deformity correction on provided images. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1030489-2020-00327
MDR Report Key9846115
Report SourceDISTRIBUTOR
Date Received2020-03-18
Date of Report2020-03-18
Date Mfgr Received2020-02-21
Date Added to Maude2020-03-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTACIE ZIEMBA
Manufacturer Street1800 PYRAMID PLACE
Manufacturer CityMEMPHIS TN 38132
Manufacturer CountryUS
Manufacturer Postal38132
Manufacturer Phone9013963133
Manufacturer G1MDT SOFAMOR DANEK PUERTO RICO MFG
Manufacturer StreetBARRIO MARIANNA RD 909, KM0.4
Manufacturer CityHUMACAO PR 00792
Manufacturer CountryUS
Manufacturer Postal Code00792
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCD HORIZON SPINAL SYSTEM
Generic NameAPPLIANCE, FIXATION, SPINAL INTERLAMINAL
Product CodeKWP
Date Received2020-03-18
Model NumberNA
Catalog Number5540030
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMDT SOFAMOR DANEK PUERTO RICO MFG
Manufacturer AddressBARRIO MARIANNA RD 909, KM0.4 HUMACAO PR 00792 US 00792


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.