WAVELIGHT EX500 EXCIMER LASER 8065990713

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-03-18 for WAVELIGHT EX500 EXCIMER LASER 8065990713 manufactured by Wavelight Gmbh.

Event Text Entries

[184318231] Investigation, including root cause analysis, is in progress. A supplemental mdr will be filed as necessary in accordance with 21 cfr 803. 56 when additional reportable information becomes available. The manufacturer internal reference number is: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[184318232] A physician reported a patient with a corneal scar in the right eye five months post photo refractive keratectomy (prk). The patient had phototherapeutic keratectomy (ptk) to remove the scar and is happy and vision is good. There are multiple patients for this reported event. This report addresses the patient pkr's left eye, and other manufacturer reports will be filed for the other patients.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003288808-2020-00206
MDR Report Key9848457
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-03-18
Date of Report2020-03-18
Date of Event2019-06-27
Date Mfgr Received2020-02-26
Device Manufacturer Date2012-11-06
Date Added to Maude2020-03-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JONATHAN SCHLECH
Manufacturer Street6201 SOUTH FREEWAY MAIL STOP AB2-6
Manufacturer CityFORT WORTH TX 76134
Manufacturer CountryUS
Manufacturer Postal76134
Manufacturer Phone8175514979
Manufacturer G1WAVELIGHT GMBH
Manufacturer StreetAM WOLFSMANTEL 5
Manufacturer CityERLANGEN 91058
Manufacturer CountryGM
Manufacturer Postal Code91058
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameWAVELIGHT EX500 EXCIMER LASER
Generic NameOPHTHALMIC EXCIMER LASER SYSTEM
Product CodeLZS
Date Received2020-03-18
Model NumberNA
Catalog Number8065990713
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerWAVELIGHT GMBH
Manufacturer AddressAM WOLFSMANTEL 5 ERLANGEN 91058 GM 91058


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.