RECON SAGITTAL SAW WITH KEY FOR TRS 05.001.240

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-18 for RECON SAGITTAL SAW WITH KEY FOR TRS 05.001.240 manufactured by Depuy Synthes Products Llc.

Event Text Entries

[184651984] Device was used for treatment, not diagnosis. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate. The device was returned for evaluation. During repair, an evaluation was performed and it was determined that the reported condition was confirmed. The assignable root cause was traced to improper maintenance. If additional information should become available, a supplemental medwatch will be submitted accordingly. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[184651985] It was reported from (b)(6) that during service and evaluation, it was determined that the recon sagittal saw device gear was worn, the thread saw blade coupling was damaged and the housing was cracked. It was further determined that the device failed pretest for general condition, leakage test using bubble emission technique, check the saw blade coupling, check roundness of housing,check falling out protection (steal ring), check for unintended motion, check fitting of the lid and check oscillation frequency with frequency meter. It was noted in the service order that the device was broken on the lid emplacement. This event did not occur during surgery. There was no patient involvement. There were no reports of injuries, medical intervention or prolonged hospitalization. All available information has been disclosed. If additional information should become available, a supplemental medwatch will be submitted accordingly.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8030965-2020-02037
MDR Report Key9848980
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-18
Date of Report2020-02-17
Date of Event2020-02-17
Date Mfgr Received2020-03-10
Device Manufacturer Date2012-12-05
Date Added to Maude2020-03-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKARA DITTY-BOVARD
Manufacturer StreetEIMATTSTRASSE 3
Manufacturer CityOBERDORF 4436
Manufacturer CountrySZ
Manufacturer Postal4436
Manufacturer Phone6103142063
Manufacturer G1SYNTHES PRODUKTIONS GMBH
Manufacturer StreetHAUPTSTRASSE 24
Manufacturer CityWALDENBURG 4437
Manufacturer CountrySZ
Manufacturer Postal Code4437
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRECON SAGITTAL SAW WITH KEY FOR TRS
Generic NameMOTOR, SURGICAL INSTRUMENT, AC-POWERED
Product CodeGEY
Date Received2020-03-18
Returned To Mfg2020-02-24
Catalog Number05.001.240
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerDEPUY SYNTHES PRODUCTS LLC
Manufacturer AddressEIMATTSTRASSE 3 OBERDORF 4436 SZ 4436


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-18