OPTICROSS 6 HD 8669

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-18 for OPTICROSS 6 HD 8669 manufactured by Boston Scientific Corporation.

Event Text Entries

[183954561] Date of event: used the first date of the month of the aware date as no event date was provided.
Patient Sequence No: 1, Text Type: N, H10


[183954562] It was reported that vessel perforation occurred. Two synergy stents where placed in the left anterior descending artery (lad) and circumflex artery (cx) using double kiss crush stenting technique. Post procedure, the patient was complaining of chest pain. The physician pulled out and advanced an opticross 6 hd intravascular ultrasound (ivus)to look on the cx but the ivus was kept on hitting the proximal portion of the stent. The ivus was pushed fairly hard and the opticross buckled, it prolapsed and perforated the lad leading to extreme complications. The patient was coded and the physician placed a covered stent in the lad. The patient survived and the procedure was completed with no further patient complications reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134265-2020-02830
MDR Report Key9849068
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-18
Date of Report2020-03-18
Date of Event2020-02-01
Date Mfgr Received2020-02-24
Date Added to Maude2020-03-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAY JOHNSON
Manufacturer StreetTWO SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal55311
Manufacturer Phone7634942574
Manufacturer G1BOSTON SCIENTIFIC IRELAND LIMITED
Manufacturer StreetBALLYBRIT BUSINESS PARK
Manufacturer CityGALWAY
Manufacturer CountryEI
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOPTICROSS 6 HD
Generic NameCATHETER, ULTRASOUND, INTRAVASCULAR
Product CodeOBJ
Date Received2020-03-18
Model Number8669
Catalog Number8669
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer AddressTWO SCIMED PLACE MAPLE GROVE MN 55311 US 55311


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-18