V60 VENTILATOR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-03-18 for V60 VENTILATOR manufactured by Respironics California, Inc.

Event Text Entries

[186687422] Date of event: (b)(6) 2020. Date of report: 18mar2020.
Patient Sequence No: 1, Text Type: N, H10


[186687463] It was reported that the unit would not transition from battery power to ac power. There was no patient involvement. The manufacturer's remote service technician performed troubleshooting with the customer. The customer reported that after a battery replacement, the unit would not power on. When the battery was removed, the unit still would not power on ac power. The customer was advised to swap out the power management board and check the power supply. The customer later reported that the power management board required replacement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2031642-2020-00957
MDR Report Key9849872
Report SourceUSER FACILITY
Date Received2020-03-18
Date of Report2020-02-28
Date Mfgr Received2020-02-28
Device Manufacturer Date2013-01-24
Date Added to Maude2020-03-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. MELISSA ABBOTT
Manufacturer Street2271 COSMOS COURT
Manufacturer CityCARLSBAD CA 92011
Manufacturer CountryUS
Manufacturer Postal92011
Manufacturer Phone9093746996
Manufacturer G1PHILIPS MEDICAL SYSTEMS
Manufacturer Street3000 MINUTEMAN ROAD
Manufacturer CityANDOVER MA 01810
Manufacturer CountryUS
Manufacturer Postal Code01810
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameV60 VENTILATOR
Generic NameVENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Product CodeMNT
Date Received2020-03-18
Model NumberV60
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerRESPIRONICS CALIFORNIA, INC
Manufacturer Address2271 COSMOS COURT CARLSBAD CA 92011 US 92011


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-18

© 2026 FDA.report