Event Text Entries

[186684691] Date of event: (b)(6) 2020. Date of report: 18mar2020.
Patient Sequence No: 1, Text Type: N, H10


[186684692] The customer reported that the power status led is green with the screen blank. The customer contacted product support and requested for service repair. The customer is unsure whether or not the blower was running. The customer disconnected the battery and reconnected. The issue goes away and is not reproducible for the biomed. The customer reviewed the error codes and reports no codes. Product support advised the customer to replace the power management board. The customer reported that the unit was not in use on a patient. The biomedical engineer replaced the power management (pm) board to address the reported issue.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2031642-2020-00958
MDR Report Key9849923
Report SourceUSER FACILITY
Date Received2020-03-18
Date of Report2020-02-28
Date Mfgr Received2020-02-28
Device Manufacturer Date2010-09-03
Date Added to Maude2020-03-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. MELISSA ABBOTT
Manufacturer Street2271 COSMOS COURT
Manufacturer CityCARLSBAD CA 92011
Manufacturer CountryUS
Manufacturer Postal92011
Manufacturer Phone9093746996
Manufacturer G1PHILIPS MEDICAL SYSTEMS
Manufacturer Street3000 MINUTEMAN ROAD
Manufacturer CityANDOVER MA 01810
Manufacturer CountryUS
Manufacturer Postal Code01810
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameV60 VENTILATOR
Generic NameVENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Product CodeMNT
Date Received2020-03-18
Model NumberV60
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerRESPIRONICS CALIFORNIA, INC
Manufacturer Address2271 COSMOS COURT CARLSBAD CA 92011 US 92011


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-18

© 2026 FDA.report