N
Patient 1
ALL ANALYSES OF THE SUSPECT SAMPLE WERE ANALYZED IN THE MANUAL MODE. CHAPTER 9 - ANALYZING SAMPLES, SECTION 9.2.1 - SAMPLE TYPES AND HANDLING, CAUTIONS USERS TO ENSURE SAMPLES ARE SUFFICIENTLY MIXED PRIOR TO ANALYSIS. A DELAY IN PROCESSING AFTER MIXING CAN LEAD TO THE PRODUCTION OF ERRONEOUS RESULTS. IN THIS MODE, THE SAMPLE MUST BE MIXED BY GENTLE INVERSION TEN TIMES PRIOR TO INTRODUCING THE SAMPLE TO THE INSTRUMENT FOR ANALYSIS. WHEN SAMPLES ARE INADEQUATELY MIXED, RED BLOOD CELLS SETTLE TO THE BOTTOM OF THE SAMPLE, CAUSING FALSELY INCREASED HGB RESULTS. IT IS THE RESPONSIBILITY OF THE USER TO MIX SAMPLES APPROPRIATELY PRIOR TO PATIENT ANALYSIS. ALL ANALYSES OF THE SUSPECT SAMPLE WERE JUDGED "POSITIVE" WITH IP MESSAGES. THE SYSMEX XN-3000/XN-3100 INSTRUCTIONS FOR USE (IFU), CHAPTER 11 - CHECKING DETAILED ANALYSIS INFORMATION (DATA BROWSER), SECTION 11.6 - IP MESSAGES, DETAILS THE METHOD IN WHICH THE ANALYZER CONVEYS ITS FINDINGS. RESULTS WITHOUT AN ERROR MESSAGE ARE CATEGORIZED AS "POSITIVE" OR "NEGATIVE" BASED UPON PRESET CRITERIA, SOME OF WHICH ARE LABORATORY-DEFINED. THE SYSTEM BASES JUDGMENTS ON COMPREHENSIVE SURVEYS OF NUMERICAL DATA, PARTICLE-SIZE DISTRIBUTIONS, AND SCATTERGRAMS. FLAGS AND MESSAGES, COMMUNICATED THROUGH IP MESSAGES, INDICATE THE ANALYZER'S FINDINGS AND NOTIFY OF POSSIBLE SAMPLE SPECIFIC ABNORMALITIES. FURTHER VERIFICATION OF ACCURATE RESULTS IS RECOMMENDED PRIOR TO REPORTING TO THE CLINICIAN. WAM DATA WAS REVIEWED. IT WAS DETERMINED THE WAM RULE TRIGGERED APPROPRIATELY AND AUTO VALIDATED THE SUSPECT HGB RESULT PER USER-DEFINED CRITERIA. THE WAM RULE THAT TRIGGERED ON THE SUSPECT SAMPLE DID NOT INCLUDE HOLDING THE HGB RESULT FOR THE DELTA CHECK. USERS ARE RESPONSIBLE FOR TESTING ALL RULES TO ENSURE THE RULE WORKS AS INTENDED PRIOR TO IMPLEMENTATION. THE ANALYZER AND WAM PERFORMED AS DESIGNED.