PLEXITRON TUR IRRIGATION SET ARC4002P

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-18 for PLEXITRON TUR IRRIGATION SET ARC4002P manufactured by Baxter Healthcare Corporation.

Event Text Entries

[184051455] (b)(6). Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[184051456] It was reported that a plexitron irrigation set was leaking at the chamber. This occurred during patient infusion. There was no report of patient injury or medical intervention associated with this event. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


[188323205] The device was received for evaluation. Visual inspection did not identify any abnormalities that could have contributed to the reported condition. All components were present and assembled according to specifications. A functional under water testing was performed which revealed bubbling through the seal zone between the chamber and the tube. Additional visual inspection was made using a stereoscope which identified a hole in the seal zone between the chamber and the tube. The area around the hole was black (similar to when plastic is burned in the machine). The reported condition was verified. The cause of the condition was due to the machine during manufacturing. A nonconformance was opened to address this issue. A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1416980-2020-01524
MDR Report Key9850765
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-18
Date of Report2020-03-18
Date of Event2020-02-16
Date Mfgr Received2020-02-21
Date Added to Maude2020-03-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPHARMACIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1BAXTER HEALTHCARE - CALI
Manufacturer StreetCALLE 36 NO. 2C-22 APARTADO AERO 2446
Manufacturer CityCALI
Manufacturer CountryCO
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePLEXITRON TUR IRRIGATION SET
Generic NameSYSTEM, IRRIGATION, UROLOGICAL
Product CodeLJH
Date Received2020-03-18
Model NumberNA
Catalog NumberARC4002P
Lot NumberSE19JJ5
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE CORPORATION
Manufacturer AddressDEERFIELD IL


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-18

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.