MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-18 for PLEXITRON TUR IRRIGATION SET ARC4002P manufactured by Baxter Healthcare Corporation.
[184051455]
(b)(6). Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10
[184051456]
It was reported that a plexitron irrigation set was leaking at the chamber. This occurred during patient infusion. There was no report of patient injury or medical intervention associated with this event. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5
[188323205]
The device was received for evaluation. Visual inspection did not identify any abnormalities that could have contributed to the reported condition. All components were present and assembled according to specifications. A functional under water testing was performed which revealed bubbling through the seal zone between the chamber and the tube. Additional visual inspection was made using a stereoscope which identified a hole in the seal zone between the chamber and the tube. The area around the hole was black (similar to when plastic is burned in the machine). The reported condition was verified. The cause of the condition was due to the machine during manufacturing. A nonconformance was opened to address this issue. A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1416980-2020-01524 |
MDR Report Key | 9850765 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2020-03-18 |
Date of Report | 2020-03-18 |
Date of Event | 2020-02-16 |
Date Mfgr Received | 2020-02-21 |
Date Added to Maude | 2020-03-18 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | PHARMACIST |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Street | 25212 W. ILLINOIS ROUTE 120 |
Manufacturer City | ROUND LAKE IL 60073 |
Manufacturer Country | US |
Manufacturer Postal | 60073 |
Manufacturer Phone | 2242702068 |
Manufacturer G1 | BAXTER HEALTHCARE - CALI |
Manufacturer Street | CALLE 36 NO. 2C-22 APARTADO AERO 2446 |
Manufacturer City | CALI |
Manufacturer Country | CO |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | PLEXITRON TUR IRRIGATION SET |
Generic Name | SYSTEM, IRRIGATION, UROLOGICAL |
Product Code | LJH |
Date Received | 2020-03-18 |
Model Number | NA |
Catalog Number | ARC4002P |
Lot Number | SE19JJ5 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BAXTER HEALTHCARE CORPORATION |
Manufacturer Address | DEERFIELD IL |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-18 |