ARCTIC SUN 5000 50000000E

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,other report with the FDA on 2020-03-19 for ARCTIC SUN 5000 50000000E manufactured by Medivance, Inc. ? 1725056.

Event Text Entries

[186335961] The investigation is still in progress. Once the investigation is complete, a supplemental report will be filed.
Patient Sequence No: 1, Text Type: N, H10


[186335962] It was reported that there was an alert 02 while initiating therapy on the arctic sun device. The flow rate was 0lpm, the inlet pressure was -7. 1 psi, the circulation pump was 51% and cv was 0. The pump hours were 762 and system hours were 818. They disconnected the pads and placed the device in manual mode. The flow rate was 0lpm, the inlet was -7. 1 psi and the circulation pump was 46%. Ms&s advised to send device to biomed. Per follow up on (b)(6) 2020, the charge nurse (b)(6) stated they swapped devices to complete therapy and the device was kept in service.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1018233-2020-01948
MDR Report Key9853821
Report SourceHEALTH PROFESSIONAL,OTHER
Date Received2020-03-19
Date of Report2020-03-19
Date Mfgr Received2020-02-28
Date Added to Maude2020-03-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYONIC ANDERSON
Manufacturer Street8195 INDUSTRIAL BLVD
Manufacturer CityCOVINGTON GA 30014
Manufacturer CountryUS
Manufacturer Postal30014
Manufacturer Phone7707846100
Manufacturer G1MEDIVANCE, INC.
Manufacturer Street321 S TAYLOR AVE
Manufacturer CityLOUISVILLE CO 80027
Manufacturer CountryUS
Manufacturer Postal Code80027
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARCTIC SUN 5000
Generic NameARCTIC SUN DEVICE
Product CodeDWJ
Date Received2020-03-19
Model Number50000000E
Catalog Number50000000E
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMEDIVANCE, INC. ? 1725056
Manufacturer Address321 S TAYLOR AVE LOUISVILLE CO 80027 US 80027


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-19