CONSTELLATION VISION SYSTEM TABLETOP 8065751150

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-03-19 for CONSTELLATION VISION SYSTEM TABLETOP 8065751150 manufactured by Alcon Research, Llc - Irvine Technology Center.

Event Text Entries

[186209110] Investigation including root cause analysis is in progress. A supplemental mdr will be filed as necessary in accordance with 21 cfr 803. 56 when additional reportable information becomes available. The manufacturer internal reference number is: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[186209111] A customer reported that during surgery, a system advisory displayed and an electrical smoke smell was observed. The patient was given a bullous of fluid so that the system could be shut down. However, the system did not maintain the eye pressure. The staff went to get a back up system, but this system would not turn on. The initial system was used to complete the case following a 20 minute delay. There was no patient harm.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2028159-2020-00257
MDR Report Key9854564
Report SourceHEALTH PROFESSIONAL
Date Received2020-03-19
Date of Report2020-03-19
Date of Event2020-02-17
Date Mfgr Received2020-02-24
Device Manufacturer Date2017-06-29
Date Added to Maude2020-03-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JONATHAN SCHLECH
Manufacturer Street6201 SOUTH FREEWAY MAIL STOP AB2-6
Manufacturer CityFORT WORTH TX 76134
Manufacturer CountryUS
Manufacturer Postal76134
Manufacturer Phone8175514979
Manufacturer G1ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
Manufacturer Street15800 ALTON PARKWAY
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal Code92618
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCONSTELLATION VISION SYSTEM
Generic NameUNIT, PHACOFRAGMENTATION
Product CodeHQC
Date Received2020-03-19
Model NumberTABLETOP
Catalog Number8065751150
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
Manufacturer Address15800 ALTON PARKWAY IRVINE CA 92618 US 92618


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-19