QUANTUM TTC ESOPHAGEAL BALLOON DILATOR QD-14X8

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-19 for QUANTUM TTC ESOPHAGEAL BALLOON DILATOR QD-14X8 manufactured by Cook Endoscopy.

Event Text Entries

[184170561] Investigation evaluation: our laboratory evaluation of the product said to be involved confirmed the report. A functional test was performed and the balloon was attempted to be inflated with water using a cook ds-60cc-s dilation syringe. The balloon could not be inflated and a leakage was observed from a rupture in the balloon material. A visual examination showed a rupture from the center of the balloon material to the proximal end of the balloon. The catheter did not show any kinks or bends. No portion of the device appeared to be missing. A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis. The device history record does contain a nonconformance that could potentially be related to the complaint. The device goes through different inspections prior to leaving the facility. These inspections would have removed any nonconforming devices prior to distribution. Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting. Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis. This limits our ability to conclusively determine a cause. A possible contributing factor for leakage is if the device comes in contact with a sharp object or burr in the endoscope accessory channel. Also, the responses to additional questions indicate that it is not known what device was used to inflate the balloon. Another possible contributing factor is using a compromised inflation device to inflate the balloon. If the pressure reading of the inflation device is inaccurate, this could contribute to over inflation, possibly resulting in damage to the balloon material. Prior to distribution, all quantum ttc esophageal balloon dilators are subjected to a leak test to ensure device integrity. Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment. Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
Patient Sequence No: 1, Text Type: N, H10


[184170562] During an esophagogastroduodenoscopy (egd), the physician used a cook quantum ttc esophageal balloon dilator. The patient aspirated after they attempted to inflate the balloon. The balloon was found to be leaking. A section of the device did not remain inside the patient? S body. The patient required antibiotics and breathing treatment due to this occurrence. According to the initial reporter, the patient did not experience any adverse effects due to this occurrence. The patient did well after the breathing treatment and antibiotics.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1037905-2020-00150
MDR Report Key9855893
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-19
Date of Report2020-02-25
Date of Event2020-02-25
Date Mfgr Received2020-02-25
Device Manufacturer Date2018-12-27
Date Added to Maude2020-03-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. SCOTTIE FARIOLE
Manufacturer Street4900 BETHANIA STATION RD
Manufacturer CityWINSTON-SALEM NC 27105
Manufacturer CountryUS
Manufacturer Postal27105
Manufacturer Phone3367440157
Manufacturer G1COOK ENDOSCOPY
Manufacturer Street4900 BETHANIA STATION RD
Manufacturer CityWINSTON-SALEM NC 27105
Manufacturer CountryUS
Manufacturer Postal Code27105
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameQUANTUM TTC ESOPHAGEAL BALLOON DILATOR
Generic NameKNQ, DILATOR, ESOPHAGEAL
Product CodeKNQ
Date Received2020-03-19
Returned To Mfg2020-03-10
Catalog NumberQD-14X8
Lot NumberW4158655
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCOOK ENDOSCOPY
Manufacturer Address4900 BETHANIA STATION RD WINSTON-SALEM NC 27105 US 27105


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.