MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-19 for TRIFECTA? GT VALVE TFGT-25A manufactured by St. Jude Medical, Costa Rica Ltda (cvd).
[184149196]
An event of insufficiency, which was suspected to be due to a torn leaflet, was reported. The results of the investigation are inconclusive since the device was not returned for analysis. The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization. Based on the information received, the cause of the reported incident could not be conclusively determined.
Patient Sequence No: 1, Text Type: N, H10
[184149197]
On (b)(6) 2016, a 25mm trifecta gt valve was implanted. The annual echocardiogram (echo) performed on (b)(6) 2018 and (b)(6) 2019 showed the valve working as intended. On (b)(6) 2020, the patient presented with aortic insufficiency from an echo exam. It was suspected that the insufficiency was due to leaflet tear. A valve in valve with a 26mm sapien valve is scheduled.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3008452825-2020-00156 |
MDR Report Key | 9855958 |
Report Source | COMPANY REPRESENTATIVE,HEALTH |
Date Received | 2020-03-19 |
Date of Report | 2020-03-19 |
Date Mfgr Received | 2020-02-25 |
Device Manufacturer Date | 2015-12-08 |
Date Added to Maude | 2020-03-19 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | PAMELA YIP |
Manufacturer Street | 5050 NATHAN LANE N |
Manufacturer City | PLYMOUTH MN 55442 |
Manufacturer Country | US |
Manufacturer Postal | 55442 |
Manufacturer Phone | 6517565400 |
Manufacturer G1 | ST. JUDE MEDICAL, COSTA RICA LTDA (CVD) |
Manufacturer Street | PARQUE INDUSTRIAL, ZONA FRANCA COYOL S.A. EDIFICIO #44B, CALLE 0, AVENIDA 2, COYOL |
Manufacturer City | ALAJUELA, COSTA RICA 01897-405 |
Manufacturer Country | CS |
Manufacturer Postal Code | 01897-4050 |
Single Use | 3 |
Remedial Action | OT |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | TRIFECTA? GT VALVE |
Generic Name | HEART-VALVE, NON-ALLOGRAFT TISSUE |
Product Code | LWR |
Date Received | 2020-03-19 |
Model Number | TFGT-25A |
Catalog Number | TFGT-25A |
Lot Number | 5337018 |
Device Expiration Date | 2017-12-07 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ST. JUDE MEDICAL, COSTA RICA LTDA (CVD) |
Manufacturer Address | PARQUE INDUSTRIAL, ZONA FRANCA COYOL S.A. EDIFICIO #44B, CALLE 0, AVENIDA 2, COYOL ALAJUELA, COSTA RICA 01897-405 CS 01897-4050 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other; 2. Required No Informationntervention | 2020-03-19 |