TRIFECTA? GT VALVE TFGT-19A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-19 for TRIFECTA? GT VALVE TFGT-19A manufactured by St. Jude Medical, Costa Rica Ltda.

Event Text Entries

[184464531] Further information regarding this event has been requested. The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
Patient Sequence No: 1, Text Type: N, H10


[184464532] On (b)(6) 2020, patient (female (b)(6) years old) had an aortic valve replacement with four bypasses. The surgeon implanted a 19mm trifecta? Gt valve. Once the patient was off pump and during the echo, they noticed a severe central leak. The surgeon decided to go back on pump and replaced the valve. An inspiris 21mm valve was implanted. This led to a prolonged procedure. The physician believes that the annulus was too tight and the trifecta valve was deformed, resulting in an incomplete leaflet coaptation. The patient remained hemodynamically stable throughout the procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3008452825-2020-00135
MDR Report Key9856270
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-19
Date of Report2020-03-19
Date of Event2020-02-19
Date Mfgr Received2020-02-20
Device Manufacturer Date2018-08-12
Date Added to Maude2020-03-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAMELA YIP
Manufacturer Street5050 NATHAN LANE N
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1ST. JUDE MEDICAL, COSTA RICA LTDA
Manufacturer StreetPARQUE INDUSTRIAL, ZONA FRANCA COYOL S.A. EDIFICIO #44B, CALLE 0, AVENIDA 2, COYOL
Manufacturer CityALAJUELA, COSTA RICA 1897-4050
Manufacturer CountryCS
Manufacturer Postal Code1897-4050
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRIFECTA? GT VALVE
Generic NameHEART-VALVE, NON-ALLOGRAFT TISSUE
Product CodeLWR
Date Received2020-03-19
Model NumberTFGT-19A
Catalog NumberTFGT-19A
Lot Number6616236
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerST. JUDE MEDICAL, COSTA RICA LTDA
Manufacturer AddressPARQUE INDUSTRIAL, ZONA FRANCA COYOL S.A. EDIFICIO #44B, CALLE 0, AVENIDA 2, COYOL ALAJUELA, COSTA RICA 1897-4050 CS 1897-4050


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2020-03-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.