BD 20ML SLIP TIP SYRINGE 301032

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a other,user facility report with the FDA on 2020-03-19 for BD 20ML SLIP TIP SYRINGE 301032 manufactured by Bd Medical (bd West) Medical Surgical.

Event Text Entries

[186323990] Investigation summary: 10 samples were received. 5 samples are luer lock syringes and 5 are luer slip syringes. One photo was provided. It shows the samples received. Two different types. It may have happened that the associate at the syringe bulk packaging area mixed the samples from the other type. A device history record review was completed with zero defects found. No quality notifications were written for this batch, nor for the associated assembly batches. Investigation conclusion: complaints received for this device and reported condition will continue to be tracked and trended. Information will be captured on trend reports and monitored monthly. Our business team regularly reviews the collected data for identification of emerging trends. Root cause description: root cause_ syringe bulk packaging process. Rationale: capa not required at this time.
Patient Sequence No: 1, Text Type: N, H10


[186323991] It was reported that bd 20ml slip tip syringe came with a mix of product types in a pack. This occurred on 8318 occasions before use. The following information was provided by the initial reporter: material no. : 301032 batch no. : 9324081. It was reported that it was discovered 20cc luer-slip syringes l-05000-004a / 23p19m0301 were mixed with 20cc luer-lock syringes within the bag.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1911916-2020-00272
MDR Report Key9856859
Report SourceOTHER,USER FACILITY
Date Received2020-03-19
Date of Report2020-03-26
Date of Event2019-12-20
Date Mfgr Received2020-03-09
Device Manufacturer Date2019-11-20
Date Added to Maude2020-03-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBDX BRETT WILKO
Manufacturer Street9450 SOUTH STATE STREET
Manufacturer CitySANDY UT 84070
Manufacturer CountryUS
Manufacturer Postal84070
Manufacturer Phone8015652341
Manufacturer G1BD MEDICAL (BD WEST) MEDICAL SURGICAL
Manufacturer Street1852 10TH AVENUE
Manufacturer CityCOLUMBUS NE 68601
Manufacturer CountryUS
Manufacturer Postal Code68601
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBD 20ML SLIP TIP SYRINGE
Generic NamePISTON SYRINGE
Product CodeFMF
Date Received2020-03-19
Returned To Mfg2020-03-17
Catalog Number301032
Lot Number9324081
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBD MEDICAL (BD WEST) MEDICAL SURGICAL
Manufacturer Address1852 10TH AVENUE COLUMBUS NE 68601 US 68601


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-03-19