10 ML BD POSIFLUSH? NORMAL SALINE SYRINGE 306595

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,other,user facility report with the FDA on 2020-03-19 for 10 ML BD POSIFLUSH? NORMAL SALINE SYRINGE 306595 manufactured by Bd Medical (bd West) Medical Surgical.

Event Text Entries

[186324459] Investigation summary: no samples were received. No photos were provided. It could have happened that a jam occurred at the plunger rod assembly process inducing damages to the barrel. A device history record review was completed with zero defects found. No quality notifications were written for this batch, nor for the associated assembly batches. Investigation conclusion: complaints received for this device and reported condition will continue to be tracked and trended. Information will be captured on trend reports and monitored monthly. Our business team regularly reviews the collected data for identification of emerging trends. Root cause description: posiflush assembly line. Plunger rod assembly. It could have happened that a jam occurred at the plunger rod assembly process inducing damages to the barrel. Rationale: capa not required at this time.
Patient Sequence No: 1, Text Type: N, H10


[186324460] It was reported that 10 ml bd posiflush? Normal saline syringe was broken and leaked. This was discovered before use. The following information was provided by the initial reporter: after opening the outer package, the flush was broken and the liquid leaked out, it was not sure that this product was able to use or not.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1911916-2020-00277
MDR Report Key9857072
Report SourceFOREIGN,OTHER,USER FACILITY
Date Received2020-03-19
Date of Report2020-03-13
Date of Event2019-06-06
Date Mfgr Received2020-03-12
Device Manufacturer Date2018-02-15
Date Added to Maude2020-03-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBDX BRETT WILKO
Manufacturer Street9450 SOUTH STATE STREET
Manufacturer CitySANDY UT 84070
Manufacturer CountryUS
Manufacturer Postal84070
Manufacturer Phone8015652341
Manufacturer G1BD MEDICAL (BD WEST) MEDICAL SURGICAL
Manufacturer Street1852 10TH AVENUE
Manufacturer CityCOLUMBUS NE 68601
Manufacturer CountryUS
Manufacturer Postal Code68601
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name10 ML BD POSIFLUSH? NORMAL SALINE SYRINGE
Generic NameSYRINGE
Product CodeFMF
Date Received2020-03-19
Catalog Number306595
Lot Number8046832
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBD MEDICAL (BD WEST) MEDICAL SURGICAL
Manufacturer Address1852 10TH AVENUE COLUMBUS NE 68601 US 68601


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-03-19