BD LUER-LOK? TIP SYRINGE 309653

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,other,user facility report with the FDA on 2020-03-19 for BD LUER-LOK? TIP SYRINGE 309653 manufactured by Bd Medical (bd West) Medical Surgical.

Event Text Entries

[186324281] Date of event: unknown. The date received by manufacturer has been used for this field. Medical device expiration date: unknown. Device manufacture date: unknown. Investigation summary: two samples were received they came wrapped in a piece of paper. One has 45ml of a solution with a tip cap; the other one is empty. Both were visually inspected first with naked eye and after with a microscope; no foreign matter was observed. Complaints received for this device and reported condition will continue to be tracked and trended. Information will be captured on trend reports and monitored monthly. Our business team regularly reviews the collected data for identification of emerging trends. A device history review could not be completed as no batch number was provided. Investigation conclusion: no defects or damages were observed. Lot# is unknown. A complaint history check was not performed as the lot number is "unknown" for this complaint. Root cause description: root cause could not be determined. Rationale: capa not required at this time.
Patient Sequence No: 1, Text Type: N, H10


[186324282] It was reported that 2 bd luer-lok? Tip syringes were found with floating foreign matter in them before use, with a"film" in the back of the plunger, and a small piece of "plastic" in the barrel. The following information was provided by the initial reporter: "we are having issues with our bd leur lock syringes. Several in the last two weeks (maybe more) have been found to have 'things' floating. They have been discovered in the pharmacy, a few were discarded without reporting, but they finally notified me last week. Last week it appeared to be a small piece of? Maybe plastic, this was a 20ml. I have three that i am currently looking at, one has a plunger that has particles of something that has migrated into the barrel of the syringe this is a 10ml, one has a film on the black part of the plunger, this is a 60ml, and one more 60ml that appears to be another small piece of plastic. "
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1911916-2020-00278
MDR Report Key9857153
Report SourceFOREIGN,OTHER,USER FACILITY
Date Received2020-03-19
Date of Report2020-03-13
Date of Event2020-03-12
Date Mfgr Received2020-03-12
Date Added to Maude2020-03-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBDX BRETT WILKO
Manufacturer Street9450 SOUTH STATE STREET
Manufacturer CitySANDY UT 84070
Manufacturer CountryUS
Manufacturer Postal84070
Manufacturer Phone8015652341
Manufacturer G1BD MEDICAL (BD WEST) MEDICAL SURGICAL
Manufacturer Street1852 10TH AVENUE
Manufacturer CityCOLUMBUS NE 68601
Manufacturer CountryUS
Manufacturer Postal Code68601
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBD LUER-LOK? TIP SYRINGE
Generic NamePISTON SYRINGE
Product CodeFMF
Date Received2020-03-19
Returned To Mfg2020-03-16
Model Number309653
Catalog Number309653
Lot NumberUNKNOWN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBD MEDICAL (BD WEST) MEDICAL SURGICAL
Manufacturer Address1852 10TH AVENUE COLUMBUS NE 68601 US 68601


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-03-19