MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-03-19 for V60 VENTILATOR 1053617 manufactured by Respironics California, Inc.
[186359942]
Date of event: (b)(6) 2020. Date of report: 19mar2020.
Patient Sequence No: 1, Text Type: N, H10
[186359943]
The customer reported touchscreen not working, and will not calibrate. It is unknown if the ventilator was being used on a patient at the time that the error was discovered. The manufacturer? S field service engineer (fse) remotely confirmed the reported touchscreen will not calibrate issue. The fse remotely confirmed that the customer replaced the touch screen to address the reported problem. The unit was checked overall, run in tested, cleaned and functionally tested and no abnormality was confirmed.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2031642-2020-00971 |
MDR Report Key | 9857677 |
Report Source | USER FACILITY |
Date Received | 2020-03-19 |
Date of Report | 2020-03-02 |
Date Mfgr Received | 2020-03-02 |
Device Manufacturer Date | 2018-09-28 |
Date Added to Maude | 2020-03-19 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | BIOMEDICAL ENGINEER |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MRS. MELISSA ABBOTT |
Manufacturer Street | 2271 COSMOS COURT |
Manufacturer City | CARLSBAD CA 92011 |
Manufacturer Country | US |
Manufacturer Postal | 92011 |
Manufacturer Phone | 9093746996 |
Manufacturer G1 | PHILIPS MEDICAL SYSTEMS |
Manufacturer Street | 3000 MINUTEMAN ROAD |
Manufacturer City | ANDOVER MA 01810 |
Manufacturer Country | US |
Manufacturer Postal Code | 01810 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | V60 VENTILATOR |
Generic Name | VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE |
Product Code | MNT |
Date Received | 2020-03-19 |
Model Number | V60 |
Catalog Number | 1053617 |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | RESPIRONICS CALIFORNIA, INC |
Manufacturer Address | 2271 COSMOS COURT CARLSBAD CA 92011 US 92011 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-03-19 |