BD SAFECLIP? 328235

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,foreign,other report with the FDA on 2020-03-19 for BD SAFECLIP? 328235 manufactured by Becton Dickinson.

Event Text Entries

[186282547] For oem manufacturing sites: in this mdr, bd corporate headquarters in (b)(4) has been listed as (b)(4) is an oem manufacturing site. Date of event: unknown. The date received by manufacturer has been used for this field. Medical device expiration date: n/a. Investigation summary: a complaint history check was performed and this is the 8th related complaint for not clipping (cutter blocked) on lot # 7177532. Customer returned a photo of a bd safe clip from lot # 7177532. Customer states that at the moment of introducing the needle to cut, it does not get into the hole, it looks like it is sealed, and sometimes the needle bends preventing the cut. The attached photo was examined and did not exhibit the cutting hole or show the ability of the sample to function. It is difficult to determine if the sample shown is able to function properly solely from the attached photo. According with the dhr review information above, the problem? No clipping? ,? Cutter blocked? And? Difficult or unable to operate? Has no nhb assembly process relation, all samples of safe clip disposable cutter (usa) passed functional test (sample plan c=0, aql=1). Based on the samples and/or photo(s) received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer? S indicated failure. Complaints received for this device and reported condition will continue to be tracked and trended. Information will be captured on trend reports and monitored monthly. Our business team regularly reviews the collected data for identification of emerging trends. Based on the above, no additional investigation and no capa is required at this time.
Patient Sequence No: 1, Text Type: N, H10


[186282548] It was reported that the safe clip is jammed with an bd safeclip?. The following information was provided by the initial reporter, translated from spanish to english: patient's mother reports that for 20 to 30 days (there is no exact date), she has been having troubles with the safe-clip. At the moment of introducing the needle to cut, it does not get into the hole, it looks like it is sealed, sometimes the needle bends preventing the cut.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2243072-2020-00454
MDR Report Key9857745
Report SourceCONSUMER,FOREIGN,OTHER
Date Received2020-03-19
Date of Report2020-03-13
Date of Event2020-03-12
Date Mfgr Received2020-03-12
Device Manufacturer Date2017-07-14
Date Added to Maude2020-03-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBDX BRETT WILKO
Manufacturer Street9450 SOUTH STATE STREET
Manufacturer CitySANDY UT 84070
Manufacturer CountryUS
Manufacturer Postal84070
Manufacturer Phone8015652341
Manufacturer G1BECTON DICKINSON
Manufacturer Street1 BECTON DRIVE
Manufacturer CityFRANKLIN LAKES NJ 07417
Manufacturer CountryUS
Manufacturer Postal Code07417
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBD SAFECLIP?
Generic NameHYPODERMIC SINGLE LUMEN NEEDLE
Product CodeFMI
Date Received2020-03-19
Model Number328235
Catalog Number328235
Lot Number7177532
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBECTON DICKINSON
Manufacturer Address1 BECTON DRIVE FRANKLIN LAKES NJ 07417 US 07417


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-03-19

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.