BROCKENBROUGH? NEEDLE EP003994S

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-19 for BROCKENBROUGH? NEEDLE EP003994S manufactured by Medtronic Mexico.

Event Text Entries

[188900902] Product event summary: the transseptal needle, ep003994s with lot number 208171403, was returned and analyzed. An image showed evidence of skiving. It is plausible that skiving occurred as there is a compatibility issue with the needle and a competitor sheath. The cause of the issue was not established. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[188900923] It was reported that during preparation of the sheath and the needle outside the patient, the user noticed that plastic material was scratched from the inside of the sheath by the needle (with dilator inserted). No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9612164-2020-01238
MDR Report Key9857940
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-19
Date of Report2020-03-19
Date of Event2014-07-17
Date Mfgr Received2014-08-11
Device Manufacturer Date2014-03-25
Date Added to Maude2020-03-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1MEDTRONIC MEXICO
Manufacturer StreetAV. PASEO DEL CUCAPAH #10510
Manufacturer CityTIJUANA,BC 22570
Manufacturer CountryMX
Manufacturer Postal Code22570
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBROCKENBROUGH? NEEDLE
Generic NameTROCAR
Product CodeDRC
Date Received2020-03-19
Returned To Mfg2014-08-21
Model NumberEP003994S
Catalog NumberEP003994S
Lot Number208171403
Device Expiration Date2016-03-23
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC MEXICO
Manufacturer AddressAV. PASEO DEL CUCAPAH #10510 TIJUANA,BC 22570 MX 22570


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-19