HEALON 5 PRO TV60ML 10240015

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-03-20 for HEALON 5 PRO TV60ML 10240015 manufactured by Johnson & Johnson Surgical Vision, Inc..

Event Text Entries

[188720936] Additional information: gender/sex: unknown, information not provided. Date of event: unknown, information not provided. If implanted, give date: not applicable as this is not an implantable device. If explanted, give date: not applicable as this is not an explantable device. Device evaluation: product testing could not be performed as the product was not returned. The reported complaint could not be verified. Manufacturing record review: the manufacturing records for the product were reviewed. The product was manufactured and released according to specifications. A search revealed that no complaints have previously been reported on this batch. Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified. All pertinent information available to johnson & johnson surgical vision, inc. Has been submitted.
Patient Sequence No: 1, Text Type: N, H10


[188720937] It was reported that the customer used healon 5 pro in two patients that experienced pressure spikes of 45 at same day postop. Thru follow-up it was reported that the operative eye is the right eye. Patient was taking daimox but discontinued taking it as the pressure improved five days post-op. Patient continues to take prednisone and ketorolac and ofloxain twice a day. Customer also reported that the patient is doing fine as the pressure went down post-op. No further information is available. This report captures patient #1. A separate medwatch is being submitted for patient #2.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004750704-2020-00021
MDR Report Key9858668
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-03-20
Date of Report2020-03-20
Report Date2005-01-01
Date Reported to FDA2005-01-01
Date Reported to Mfgr2005-01-10
Date Mfgr Received2020-02-21
Date Added to Maude2020-03-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSOMYATA NAGPAL
Manufacturer Street1700 EAST ST. ANDREW PLACE
Manufacturer CitySANTA ANA CA 92705
Manufacturer CountryUS
Manufacturer Postal92705
Manufacturer Phone7142478200
Manufacturer G1JOHNSON & JOHNSON SURGICAL VISION, INC.
Manufacturer StreetRAPSGATAN
Manufacturer CityUPPSALA 751 82
Manufacturer CountrySE
Manufacturer Postal Code751 82
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameHEALON 5 PRO
Generic NameOVDS
Product CodeLZP
Date Received2020-03-20
Model NumberTV60ML
Catalog Number10240015
Lot NumberUE31141
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Sequence No1
Device Event Key0
ManufacturerJOHNSON & JOHNSON SURGICAL VISION, INC.
Manufacturer Address1700 E ST ANDREW PLACE SANTA ANA CA 92705 US 92705

Device Sequence Number: 101

Product Code---
Date Received2020-03-20
Device Sequence No101
Device Event Key0


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-03-20

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.