TARGET TIP 506158

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-03-20 for TARGET TIP 506158 manufactured by Medtronic, Inc..

Event Text Entries

[184301376] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[184301377] It was reported that capture management on right ventricular (rv) lead was unstable. It was also reported that the rv lead exhibited experienced oversensing with an elevated sensing integrity counter (sic). The rv lead remains in use. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2182208-2020-00535
MDR Report Key9859614
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-03-20
Date of Report2020-03-20
Date of Event2020-02-25
Date Mfgr Received2020-02-25
Date Added to Maude2020-03-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1MEDTRONIC, INC.
Manufacturer Street8200 CORAL SEA STREET NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal Code55112
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTARGET TIP
Generic NameADAPTOR, LEAD, PACEMAKER
Product CodeDTD
Date Received2020-03-20
Model Number506158
Catalog Number506158
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC, INC.
Manufacturer Address8200 CORAL SEA STREET NE MOUNDS VIEW MN 55112 US 55112


Patients

Patient NumberTreatmentOutcomeDate
15076 2020-03-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.