18IN. (46CM) A.T.S. CYLINDRICAL CUFF - DP, DB 60755500200

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-03-20 for 18IN. (46CM) A.T.S. CYLINDRICAL CUFF - DP, DB 60755500200 manufactured by Zimmer Surgical, Inc..

Event Text Entries

[186712016] This event has been recorded under zimmer biomet complaint number (b)(4). The customer has indicated that the device will be returned for evaluation and investigation is in process. Once the investigation is complete, a supplemental report will be filed accordingly.
Patient Sequence No: 1, Text Type: N, H10


[186712017] It was reported that the clinic tested the tourniquet cuff before using it. They found that once it inflates they could not deflate it without detaching the hoses. It was not used for any procedure. No adverse events were reported as a result of this malfunction.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0001526350-2020-00298
MDR Report Key9859888
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-20
Date of Report2020-03-20
Date of Event2020-02-19
Date Mfgr Received2020-02-24
Date Added to Maude2020-03-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. CHRISTINA ARNT
Manufacturer Street56 E. BELL DRIVE
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal46582
Manufacturer Phone5745273773
Manufacturer G1ZIMMER SURGICAL, INC.
Manufacturer Street200 WEST OHIO AVENUE
Manufacturer CityDOVER OH 44622
Manufacturer CountryUS
Manufacturer Postal Code44622
Single Use3
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand Name18IN. (46CM) A.T.S. CYLINDRICAL CUFF - DP, DB
Generic NameTOURNIQUET CUFF, REUSABLE
Product CodeKCY
Date Received2020-03-20
Catalog Number60755500200
Lot NumberNI
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerZIMMER SURGICAL, INC.
Manufacturer Address200 WEST OHIO AVENUE DOVER OH 44622 US 44622


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-20

© 2026 FDA.report