[184464719] One brk transseptal needle was received for evaluation. The brk needle had been returned coiled and was kinked in multiple locations along its length. The needle tip was fractured and detached. Functional testing was not possible due to the aforementioned damage. The device history record was reviewed to ensure that each manufacturing and inspection operation was performed. Based on the received condition of the device and the information provided, the cause of reported event remains unknown.
Patient Sequence No: 1, Text Type: N, H10


[184464720] This is event is being reported based on analysis revealing a detached needle tip.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3008452825-2020-00155
MDR Report Key9860039
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-20
Date of Report2020-03-20
Date of Event2019-12-26
Date Mfgr Received2020-03-10
Device Manufacturer Date2019-07-23
Date Added to Maude2020-03-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEPHANIE O' SULLIVAN
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1ST. JUDE MEDICAL, COSTA RICA LTDA
Manufacturer StreetPARQUE INDUSTRIAL, ZONA FRANCA COYOL S.A. EDIFICIO #44B, CALLE 0, AVENIDA 2, COYOL
Manufacturer CityALAJUELA, COSTA RICA 1897-4050
Manufacturer CountryCS
Manufacturer Postal Code1897-4050
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBRK? TRANSSEPTAL NEEDLE, 71 CM LENGTH
Generic NameTROCAR
Product CodeDRC
Date Received2020-03-20
Model Number407200
Catalog Number407200
Lot Number7101117
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerST. JUDE MEDICAL, COSTA RICA LTDA
Manufacturer AddressPARQUE INDUSTRIAL, ZONA FRANCA COYOL S.A. EDIFICIO #44B, CALLE 0, AVENIDA 2, COYOL ALAJUELA, COSTA RICA 1897-4050 CS 1897-4050


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.