[185674023] Patient was reported to have no adverse reaction for this event; however, lost 10 liters of blood during the procedure. A loss of 3. 5 liters of blood was due to vaginal bleeding that could not be salvaged. A field service engineer was dispatched and inspected the machine and replaced a pinch valve bushing. Diagnostics checks were run and unit meets manufacturers' specifications. The disposable processing kit that was used in this event was discarded. This event occurred in the united kingdom with a product that is not registered or marketed within the us. However, haemonetics has a similar device marketed and cleared for use by the fda where the alleged malfunction can potentially occur. This report is being filed due to the reported failures that could occur on a similar device marketed and cleared for use by the fda.
Patient Sequence No: 1, Text Type: N, H10


[185674024] On (b)(6) 2020, haemonetics was informed by the customer of an adverse event and device malfunction and adverse event that happened during a gynecological procedure. A long empty alarm was sounded and a second machine was brought in to finish the procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1219343-2020-00020
MDR Report Key9860216
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-03-20
Date of Report2020-02-19
Date of Event2020-02-19
Date Mfgr Received2020-02-19
Date Added to Maude2020-03-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR SHAUN FLANAGAN
Manufacturer Street125 SUMMER STREET
Manufacturer CityBOSTON, MA
Manufacturer CountryUS
Manufacturer G1HAEMONETICS CORPORATION
Manufacturer Street125 SUMMER STREET
Manufacturer CityBOSTON, MA
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCELL SAVER ELITE+, UK
Generic NameCSE-EW-UK
Product CodeCAC
Date Received2020-03-20
Model NumberCELL SAVER ELITE+, UK
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHAEMONETICS CORPORATION
Manufacturer Address125 SUMMER STREET BOSTON, MA US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-03-20