COAGUCHEK ? XS SYSTEM 03666794001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2020-03-20 for COAGUCHEK ? XS SYSTEM 03666794001 manufactured by Roche Diagnostics.

Event Text Entries

[184302851] The customer? S meter was requested for return. The investigation is ongoing.
Patient Sequence No: 1, Text Type: N, H10


[184302852] The initial reporter complained of a burning or charring issue with coaguchek xs meter serial number (b)(4). The customer complained of various error messages (example: error 4) and was unable to perform a test. It was noted that the heater plate appeared to be charred or burnt. There were no other signs of burning, melting, or smoking to any other portion of the meter.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2020-00793
MDR Report Key9860755
Report SourceCONSUMER
Date Received2020-03-20
Date of Report2020-03-20
Date of Event2020-02-24
Date Mfgr Received2020-02-24
Date Added to Maude2020-03-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOAGUCHEK ? XS SYSTEM
Generic NamePROTHROMBIN TIME MONITOR
Product CodeGJS
Date Received2020-03-20
Model NumberNA
Catalog Number03666794001
Lot NumberNA
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2020-03-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.