PCS?2 PLASMA COLLECTION SYSTEM 06002-CP-110

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-03-20 for PCS?2 PLASMA COLLECTION SYSTEM 06002-CP-110 manufactured by Haemonetics Corporation.

Event Text Entries

[184303511] Donor was monitored and was stable upon release. A field service engineer was dispatched and inspected the pcs2 machine. The engineer realigned centrifuge lid and tightened loose hardware. The blood pump motor was also replaced. Calibrations checked and adjusted as required. Unit meets manufacturer specifications. Photo analysis of the bowl shows a bowl full of dark colored red blood cells. No evidence of damaged red blood cells are visible. No defects that could possibly create hemolysis were identified (no folds, molding defects, sharp or rough edges, foreign objects).
Patient Sequence No: 1, Text Type: N, H10


[184303512] On february 21, 2020, haemonetics was informed by the customer of an adverse event and device malfunction during a plasmapheresis procedure. There was a possibility of the donor receiving hemolized red blood cells.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1219343-2020-00024
MDR Report Key9860898
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-03-20
Date of Report2020-02-21
Date of Event2020-02-21
Date Added to Maude2020-03-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR SHAUN FLANAGAN
Manufacturer Street125 SUMMER STREET
Manufacturer CityBOSTON, MA
Manufacturer CountryUS
Manufacturer G1HAEMONETICS CORPORATION
Manufacturer Street125 SUMMER STREET
Manufacturer CityBOSTON, MA
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePCS?2 PLASMA COLLECTION SYSTEM
Generic NamePCS2,LIST NO. 06002-CP-110
Product CodeGKT
Date Received2020-03-20
Model Number06002-CP-110
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHAEMONETICS CORPORATION
Manufacturer Address125 SUMMER STREET BOSTON, MA US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-03-20

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.