NEEDLE ECLIPSE 23X1 RB 305762

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a other,user facility report with the FDA on 2020-03-20 for NEEDLE ECLIPSE 23X1 RB 305762 manufactured by Becton Dickinson Medical (singapore).

Event Text Entries

[188220753] Device expiration date: unknown. The reported lot# 9267472 was not found for the catalog number. Device manufacture date: unknown. Investigation summary: no photo/ sample were returned for investigation. Root cause could not be determined. The complaint will be re-opened and re-investigated if the sample is returned. Investigation conclusion: unable to confirm the customer experience as no photo/ sample was returned for investigation. Root cause description: no photo/ sample were returned for investigation. Root cause could not be determined.
Patient Sequence No: 1, Text Type: N, H10


[188220754] It was reported that a safety mechanism failure occurred during use with a needle eclipse 23x1 rb. The following information was provided by the initial reporter, "the safety guard on the needle had broken when the nurse attempted to push the safety guard down over the needle, resulting in the stick. The plastic of the safety guard broke at the base of the needle where the guard hinges with the needle. " no medical intervention was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8041187-2020-00158
MDR Report Key9861834
Report SourceOTHER,USER FACILITY
Date Received2020-03-20
Date of Report2020-03-03
Date of Event2020-02-26
Date Mfgr Received2020-03-02
Date Added to Maude2020-03-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBDX BRETT WILKO
Manufacturer Street9450 SOUTH STATE STREET
Manufacturer CitySANDY UT 84070
Manufacturer CountryUS
Manufacturer Postal84070
Manufacturer Phone8015652341
Manufacturer G1BECTON DICKINSON MEDICAL (SINGAPORE)
Manufacturer Street30 TUAS AVENUE 2
Manufacturer CitySINGAPORE
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNEEDLE ECLIPSE 23X1 RB
Generic NameHYPODERMIC SINGLE LUMEN NEEDLE
Product CodeFMI
Date Received2020-03-20
Model Number305762
Catalog Number305762
Lot NumberSEE. H.10
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBECTON DICKINSON MEDICAL (SINGAPORE)
Manufacturer Address30 TUAS AVENUE 2 SINGAPORE


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-03-20

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.